BuffermetryFundamentals of moisture buffering Properties of hygroscopic materials Diffusion enhanced materials and structures Experimental: Sorption experiments Experimental: Diffusion experiments Incorporating the buffer in the construction Whole room simulations and experiments Experiment in the Passys environmental room Appendices - instruments and procedures Wiki howto |
Tinman maintenance notesTo close down the instrument safely: From the terminal running climchxxx.py type control+c. This will interrupt the program so it closes down safely. If you are in a different terminal (meaning you cannot see the program output to screen) type: This will show all the processes. Find the process number for python climch.... which is, for example 3572 type: This will NOT close the program safely, so you should then type: wait for it to return a prompt: your command... then type x This will close down the instrument safely. In the chamber: Before doing maintenance which involves much restarting of the program, disconnect the peltier leads at the two-connector plug (the peltier leads are the thick red and red/black leads). Before re-connecting, test for high resistance (at least 1 Mohm) between the peltier terminals on the weighing device and the metal parts of the system. Then test for a fairly low resistance, maybe 15 ohm, across the peltier terminals at the end of the thin steel conductor strips. |